Question: How do I know I am getting the correct results with SystemSURE II and Ultrasnap?

Answer:
The SystemSURE II is programmed to self-calibrate every time the instrument is turned on. During the 60-second calibration, it is measuring the temperature and humidity, as well as checking the light sensor. If you are unsure whether your instrument is calibrating properly, you can run a positive and negative control test using a “Calibration Control Kit” (Part #: PCD4000). You can also check the sample testing devices with the Positive Control Kit - (Part #:PC25).